리퍼비시 계측기
The Keysight (Agilent/HP) 8902A measuring receiver delivers the accuracy and resolution of a high performance power meter at frequencies from 150 kHz to 1.3 GHz (50 MHz to 26.5 GHz with the Keysight (Agilent/HP) 11793A microwave converter) and levels from +30 dBm to -127 dBm. It accurately measures AM, FM and M, including residuals and incidentals, with a single keystroke. The Keysight (Agilent/HP) 8902A measuring receiver, with the Keysight (Agilent/HP) 11793A, counts RF signals to 26.5 GHz with 10 Hz resolution and excellent long-term frequency stability. The 8902A measuring receiver with Option 050 offers increased power measurement accuracy. This option specifies Tuned RF Level on the 8902A measuring receiver to an accuracy of +/-(0.015 dB + 0.005 dB/10 dB step).
Frequency Modulation
RATES:
20 Hz to 10 kHz, 150 kHz ≤ fc <10 MHz.
20 Hz to 200 kHz, 10 MHz ≤ fc ≤1300 MHz.
20 Hz to 200 kHz, 10 MHz ≤ fc ≤26.5 GHz.
DEVIATIONS:
40 kHz peak maximum, 150 kHz ≤ fc <10 MHz.
400 kHz peak maximum, 10 MHz ≤ fc ≤1300 MHz.
400 kHz peak maximum, 10 MHz ≤ fc ≤26.5 GHz.
ACCURACY :
FM Accuracy /Frequency Range /Rates Deviations
±2% of reading / 250 kHz – 10 MHz / 20 Hz – 10 kHz / ≤40 kHz peak
±1 digit
±1% of reading / 10 MHz – 1300 MHz / 50 Hz – 100 kHz / ≤400 kHz peak
±1 digit
±5% of reading / 10 MHz – 1300 MHz / 20 Hz – 200 kHz / ≤400 kHz peak
±1 digit
±1% of reading / 10 MHz – 26.5 GHz / 50 Hz – 100 kHz / ≤400 kHz peak
±1 digit
±5% of reading /10 MHz – 26.5 GHz / 20 Hz – 200 kHz / ≤400 kHz peak
±1 digit
DEMODULATED OUTPUT DISTORTION :
THD / Frequency Range /Rates / Deviations
<0.1% / 400 kHz – 10 MHz / 20 Hz – 10 kHz / <10 kHz
<0.1% / 10 MHz – 1300 MHz / 20 Hz – 100 kHz / <100 kHz
<0.1% / 10 MHz – 26.5 GHz / 20 Hz – 100 kHz / <100 kHz