리퍼비시 계측기
The Hioki 8960 Strain Unit is an optional product for Hioki Memory HiCorder devices. Be sure to use this input module only by installing it in a Memory HiCorder. For information on how to install and use the input module, refer to the documentation for the Memory HiCorder.
Number of input channels | 2 channels |
Input terminals | R05-R5F by TAJIMI |
Appropriate adapter | Strain gauge adapter |
Bridge voltage | 2, 5, 10 ± 0.05 V |
Bridge resistance | 120 Ω to 1 kΩ (Bridge voltage 2 V) 350 Ω to 1 kΩ (Bridge voltage 5, 10 V) |
Balance adjustment range | ±10000 με max |
Balancing | Electronic auto-balancing |
Measurement ranges | 20, 50, 100, 200, 500, 1000 με/div |
DC amplitude accuracy | ±(0.4%f.s.+ 2 με) (filter 5 Hz ON) (When using the conversion cables supplied. When using other cables, cable resistance is 1.5 Ω max. at bridge resistance 350 Ω) |
Zero position accuracy | ±(0.1%f.s.+ 2 με) (filter 5 Hz ON, after auto-balancing) |
Temperature characteristic | Gain: ±0.05%f.s./°C Zero position (after auto-balancing):±2.5 με/°C |
Frequency characteristic | DC to 20 kHz dB |
Low-pass filter | OFF, 5±30%, 10±30%, 100±30%, 1k±30% (Hz) -3 dB |
Anti-aliasing filter | Cutoff frequency (fc) 20, 40, 80, 200, 400, 800, 2 k, 4 k, 8 k, 20 k, 40 k (Hz) (These frequencies are automatically set when the anti-aliasing filter is set to ON) Attenuation; -66 dB min at 1.5 fc A/D resolution 16 bits |
Maximum sampling rate | 200 kS/s |
Maximum input voltage | 10 VDC |
Maximum rated voltage to earth | 33 Vrms or 70 VDC (between each input channel and main unit, and between input channels) |
Dielectric strength | 350 VAC for 15 seconds (between input module and main unit, and between input modules) |
Operating temperature and humidity ranges | Same as the Memory HiCorder in which the 8960 is installed |
Operating environment | Same as the Memory HiCorder in which the 8960 is installed |
Storage temperature and humidity ranges | Same as the Memory HiCorder in which the 8960 is instal |