리퍼비시 계측기
■ Generator specifications
Parameter | Specification / Function |
---|---|
Output voltage | 200 to 5000 V 10 V Step |
Polarity | Positive or negative, polarity alternation possible per burst |
Repetition frequency | 0.1 kHz to 2000 kHz 0.1 kHz to 1 kHz / 0.01 kHz step Tolerance ± 5%, 1.0 kHz to 10 kHz / 0.1 kHz step Tolerance ± 5% 10 kHz to 100 kHz / 1 kHz step Tolerance ± 5%, 100 kHz to 1000 kHz / 10 kHz step Tolerance ± 5% 1000 kHz to 2000 kHz / 100 kHz step Tolerance ± 10%, (Limitation per voltage levels when continuous output) |
Number of pulses | 1 to 1000 at a step of 1 pulse , Setting limit: 1 pulse per ms in a burst (repetition frequency 1 kHz or more) |
Burst duration | Formula for Burst duration = Pulse number / Repetition Frequency Scope of manually setting value for burst duration: 0.01 to 999 ms |
Burst period | 10 to 1000ms ± 10% 10ms steps (500ms or more for polarity alternate mode) |
Polarity alternate function | Output polarity alternated between positive and negative at each burst period Setting condition: the burst period is 500ms or more and the burst pause period [ (burst period) – (burst duration)] is 100ms or more Maximum test time: 10 minutes |
Continuous Pulse output | Up to 1000 V -10 kHz or less, to 2000 V -4 kHz or less, to 5000 V -1 kHz or less. Maximum test time for each case: 10 min |
Frequency modulation | Frequency is shifted continuously between set frequency and approximately -10% from the set frequency. The modulating wave is triangular wave of approximately 20Hz |
External trigger | External trigger input invokes 1 burst output in synchronization with the trigger input. Trigger specification: Hi (+ 5V) → Lo (0 V) triggers one burst period. |
Pulse waveform (at 50 Ω load) | Pulse peak voltage: (set voltage / 2) ± 10% Rise time: 5 ns ± 30% Pulse width: 50 ns ± 30% |
Pulse waveform (at 1 kΩ load) | Pulse peak voltage: (set voltage x 0.95) ± 20% Rise time: 5 ns ± 30% Pulse width: 35 to 150 ns |
DC blocking capacitor | 10nF ± 20% |
■ CDN specifications
Parameter | Specification / Function |
---|---|
Power capacity | A 20 model: single phase AC 240 V / 20 A, DC 125 V / 20 A (10 A for PE) B 63 model: three-phase AC 600 V / 63 A, DC 125 V / 63 A (10 A for N / PE) |
Applied phase | A20 model: L / N / PE B63 model: L1 / L2 / L3 / N / PE Single line or all lines can be specified individually for each phase |
Injection mode | Common mode (Normal mode available using option) |
EUT Line input/output | φ6 mm safety socket |
Coupling capacitor | 33 nF |
Output waveform specification | Pulse peak voltage: (set voltage) / 2 ± 10% Rise time: 5.5 ns ± 1.5 ns Pulse width: 45 ns ± 15 ns Set voltage ± 4000 V, frequency specified from 5 kHz to 100 kHz |
Input residual voltage | 10% or less of setting pulse voltage EUT line input is 50 Ω termination, line output is defined as open |
AC Line Sync | Synchronous and asynchronous setting available. Setting phase angle: 0 to 360 ° ± 10 ° 1 °Step Synchronizable voltage: AC 85 V to rated voltage Reference phase: between L-N (A20 model), L1-L2 (B63 model) |
■ Other specifications
Parameter | Specification / Function |
---|---|
Emergency stop | Push lock type switch (Test stop, EUT line OFF) |
EUT FAIL function | FAIL signal from external (Hi → Lo) detected during test FAIL signal specifications VLO: 0 V, VHI: + 5 V Choose operation from test stop / pause when triggered 3 channels available for the FAIL input |
External interface | REMOTE (For external PC control), CDN I/F (For external CDN), INDICATOR (For Warning Lamp or indicator lamp) EUT FAIL INPUT (For temporary pause at EUT failure event) |
Accessories | Power Cable, SG Cable, Line Input Cable, Output Cable, Waveform Check Connector, Coaxial Cable, Operation Manual, Accessoriesies bag |
Operating environment | Temperature 15 to 35 °C Relative humidity 25 to 75% |
External Dimensions / Weight | (W)430 x (H)199 x (D)370 mm (excluding protrusions) / Approximately 14 kg (A20 model) and 22 kg (B63 model) |
Power supply | AC 100 to 240 V ± 10% 50/60 Hz approx. 120 VA |